Model Number G8650 |
Device Problems
Break (1069); Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/16/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The incident device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
|
|
Event Description
|
Procedure performed: unspecified cardiovascular surgery.Incident description: "geomed clamp was used to attach the insert g8650.A medical staff tried to clamp the patient's vessel.At that time, the medical staff found that the insert wobbled a little, but the medical staff elected to use the insert.When the medical staff tried to remove the clamp from the patient vessel, the medical staff found that one of the attachment button of the insert was missing.The medical staff thought that the attachment button was fallen into the patient body.At the end, the missing attachment button was recovered by cell saver® (blood recovery system).This event didn't affect the patient." type of intervention: na.Patient status: "no health damage was reported for the patient.".
|
|
Manufacturer Narrative
|
The event unit returned to applied medical for evaluation.Upon visual inspection, engineering confirmed that the distal button was missing from the fibra insert.Additional damage was observed on both the soft insert and fibra insert.However, the damage was caused by the complainant and was unrelated to the reported event.Based on the condition of the returned unit, it is likely that the reported event was caused when the insert was removed from the clamp.The instructions for use (ifu) indicates, "[t]o remove inserts, open the clamp and hold the tip of the insert between the bottom edge and the clamp jaw.Pull outward to remove the insert.".Applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.
|
|
Event Description
|
Additional information received via email from regulatory affairs and quality assurance division, on 08aug2017 - the hospital has used the geomed clamp for more than 2 years and no complaint was reported until this time.Specifically, the hospital has used amr insert with the clamp before.There was no detailed information on the geomed clamp such as photos and model number.For the specific clamp used during the procedure, the clamp is in good condition.There is no further details on the sequence of events with regards to the insert button being seen as missing.The inserts was fractured intentionally by medical staff.No related to this event (button missing).
|
|
Search Alerts/Recalls
|
|