• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590 BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER Back to Search Results
Model Number 119116
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the tip of the catheter and the inflation lumen areas appeared to have excess silicone.This complaint was opened because associate fa engineer found issues outside of what was originally reported.
 
Additional Manufacturer Narrative
Received 1 used drainage bag with the catheter still attached and the syringe.The reported event was confirmed as manufacturing related.Per the visual inspection, a burr was noted at the catheter tip and the funnel.No other defects were found.Per the dimensional evaluation, the flash on molded funnel was found outside the specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.".
 
Event or Problem Description
It was reported that the tip of the catheter and the inflation lumen areas appeared to have excess silicone.This complaint was opened because associate fa engineer found issues outside of what was originally reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE TEMPERATURE-SENSING FOLEY CATHETER
Common Device Name
CATHETER
Manufacturer (Section D)
PRODUCTOS PARA EL CUIDADO DE LA SALUD -9611590
km. 7 carretera internacional
nogales, sonora 85621
MX  85621
MDR Report Key6672506
Report Number1018233-2017-03307
Device Sequence Number2059429
Product Code EZL
UDI-Device Identifier00801741046032
UDI-Public(01)00801741046032
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model Number119116
Device Catalogue Number119116
Device Lot NumberNGBP2012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2017
Initial Date Received by Manufacturer 06/05/2017
Supplement Date Received by Manufacturer07/31/2017
Initial Report FDA Received Date06/27/2017
Supplement Report FDA Received Date08/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-