| Model Number |
382401 |
| Medical Device Problem Code |
Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Submit date: 2017-06-28.
|
| |
|
Event or Problem Description
|
|
The customer unit was returned for routine maintenance.Upon triage, the unit shut off and had intermittent power during the recertification test.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the kangaroo pump was performed for the reported condition of shut off and had intermittent power during recertification test.The unit was triaged and the reported issue was confirmed.The pump was excessively damaged, so the root cause is categorized as customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
| |
|