• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL SURGICAL INNOVATIONS -9616067 MAGIC3 TOUCHLESS¿ HYDROPHILIC INTERMITTENT CATHETER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL SURGICAL INNOVATIONS -9616067 MAGIC3 TOUCHLESS¿ HYDROPHILIC INTERMITTENT CATHETER SYSTEM Back to Search Results
Model Number 58814
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the once the water sachet was burst it made the hydrophilic coating very thick and cake like.The consistency of the coating clogged the catheter.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use state the following: "instructions for use: preparing for catheterization: open kit and remove contents.Wash hands.Activate the catheter¿s hydrophilic coating by releasing the sterile water from the foil packet into the collection bag.Apply pressure to the foil packet to release the water.Ensure all water is released from the foil packet.Wet the catheter coating by tipping the collection bag end-to-end six or more times.This transfers the enclosed water back and forth over the catheter to fully wet the hydrophilic coating.Ensure water flows into the insertion tip to fully wet the catheter tip.Male/female catheterization: male - prepare the male urethral meatus and the surrounding area with the antiseptic provided.Female - prepare the female urethral meatus by holding the outer labia apart and prep the urethral meatus and surrounding area with the antiseptic provided.Remove white cap from insertion tip while squeezing tabs.Save cap for closing the bag.Slide the catheter halfway into the insertion tip.Male - holding the penis, advance the insertion tip into the urethra no further than the flange base.Female - spread the inner labia, advance the insertion tip into the urethra no further than the flange base.Release the inner labia.Place your nondominant hand on the sure-grip sleeve to stabilize catheter in urethra.With your dominant hand, grasp catheter through bag approximately 1" below the sure-grip sleeve and push catheter into urethra.The catheter should be introduced by short, repetitive pushing motions.Repeat motions until catheter reaches bladder and urine starts to flow.Allow urine to flow freely, making certain the sure-grip sleeve is elevated at least 4" above lower portion of the bag.Allow flow until bladder is empty or until bag is filled.Precaution: it is recommended that the collection bag be held.The catheter could possibly separate from the collection bag when urine increases weight of the bag.Withdraw the catheter from the urethra.If desired, the remaining portion of the catheter can be removed from the collection bag by pulling it through the insertion tip.Note: the catheter is designed to pass through the insertion tip.The filled collection bag may be closed by replacing the cap over the insertion tip.Make sure the cap snaps on securely.Specimen collection: to collect a specimen, obtain a sample cup/tube and alcohol wipe.Remove tip by pulling tab(s) upward from bag.Wipe port with alcohol.Pour specimen through port at top of bag into cup/tube.Draining bag: option 1 - remove insertion tip by pulling tab(s) upward from bag.Drain urine through port at top of bag.Option 2 - open bag by tearing at the marked perforation.Drain urine through opening." the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that once the water sachet was burst it made the hydrophilic coating very thick and cake like.The consistency of the coating clogged the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3 TOUCHLESS¿ HYDROPHILIC INTERMITTENT CATHETER SYSTEM
Common Device Name
TOUCHLESS
Manufacturer (Section D)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX  32690
MDR Report Key6675936
Report Number1018233-2017-03350
Device Sequence Number1601897
Product Code MJC
UDI-Device Identifier00801741104053
UDI-Public(01)00801741104053
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K033477
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,other
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 07/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number58814
Device Catalogue Number58814
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/07/2017
Supplement Date Received by Manufacturer07/14/2017
Initial Report FDA Received Date06/28/2017
Supplement Report FDA Received Date07/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-