• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 1125400-15
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for analysis.The reported collapse tip was unable to be confirmed; however, the stent was noted to be dislocated on the balloon.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.Due to the extent of the damage to the stent, it is unlikely that the protective sheath would have fit over the damage; therefore, the damage is not a pre-existing condition.It is likely that the reported tip collapse and noted stent dislocation was a result of inadvertent mishandling of the sds during unpackaging and/or during preparation and there is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event or Problem Description
It was reported that the 4.0 x 15 mm xience alpine was noted to have a deformed tip upon removal from the dispenser hoop.There was no patient involvement and no clinically significant delay.Returned device analysis noted that the stent was stationary on the balloon, but was not between the markers.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6676038
Report Number2024168-2017-05509
Device Sequence Number1615073
Product Code NIQ
UDI-Device Identifier08717648199615
UDI-Public(01)08717648199615(17)191005(10)6092741
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 06/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/05/2019
Device Model Number1125400-15
Device Catalogue Number1125400-15
Device Lot Number6092741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2017
Initial Report FDA Received Date06/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-