| Catalog Number |
7204 |
| Medical Device Problem Codes |
Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This is an initial and final report in one.The device has not been returned from the reporter so a thorough investigation cannot be conducted.A theoretical evaluation can however be done and our conclusion is that the device has been used beyond its expected lifetime or used without adhering to the guidelines in the ifu of inspecting prior to use, not bending the brush repetitively and to discard when worn.Brush and part discarded by user.
|
| |
|
Event or Problem Description
|
|
This is the information that was received from the atos medical local representative: the patient reports that she aspirated the metal/brush part of the tep brush while cleaning her tep.She reports it being in the lumen of the tep and when she pulled it out the metal part with the bristles fell off of the brush and into her airway.She reported that the brush was given to her by her slp ((b)(6)) approximately 2 mos.Prior to the event.She leaned over and was able to cough this piece of the brush out of her airway.None of the parts were kept to be evaluated.The slp does have the lot number and the packs of brushes she recently ordered.
|
| |
|