• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS CORPORATION - MONTGOMERY 3085SP SURGICAL TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS CORPORATION - MONTGOMERY 3085SP SURGICAL TABLE Back to Search Results
Device Problem Failure to Read Input Signal (1581)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2017
Event Type  malfunction  
Manufacturer Narrative
A procedure delay occurred due to the transfer of the patient.A steris service technician arrived onsite to inspect the table and found two wires comprising the ls4 circuit of the back section cable assembly were pinched between the back section frame and back section top.A loop of the wire circuit was projecting outside the interior of the back section frame; the wire circuit should have been secured.The service technician repaired the surgical table, tested the device and confirmed it to be operating properly.The surgical table was returned to service and no additional issues have been reported.
 
Event Description
The user facility reported that during the start of a patient procedure the table would not respond to commands via the hand control.The patient was transferred to a different table and the procedure was completed successfully.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3085SP SURGICAL TABLE
Type of Device
SURGICAL TABLE
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
lyndsey snyder
5960 heisley road
mentor, OH 44060
4403927386
MDR Report Key6677544
MDR Text Key78970616
Report Number1043572-2017-00044
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2017
Initial Date FDA Received06/29/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-