A procedure delay occurred due to the transfer of the patient.A steris service technician arrived onsite to inspect the table and found two wires comprising the ls4 circuit of the back section cable assembly were pinched between the back section frame and back section top.A loop of the wire circuit was projecting outside the interior of the back section frame; the wire circuit should have been secured.The service technician repaired the surgical table, tested the device and confirmed it to be operating properly.The surgical table was returned to service and no additional issues have been reported.
|