Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report.During our evaluation of the device, a functional test was performed.The device was tested for opening and closing of the forceps cups.When the handle was manipulated, the forceps cups would open, but they would not close.The device was then placed down an olympus gif-q20 (2.8 mm channel) endoscope.The endoscope was placed in a tortuous path.When the handle was manipulated the forceps cups were slow to open, it took several handle manipulation attempts for the forceps cups to open fully.Once the cups were open, the cups would not close.The device was sent back to the supplier.The supplier provided the following evaluation: one device from the reported event was returned in a zip type bag with proof of decontamination.Evaluation of the device determined that for the returned device the device was tested for "would not close." during functional testing, with the device coiled into three (3), approximately eight (8) inch loops, it was confirmed that the device would not operate properly when the handle was manipulated.The device opens but does not close.Upon further investigation and disassembly of the device tip, it was noted that the device has a broken solder joint.The reported defect was confirmed.Failure was due to a broken solder joint.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the forceps cups will open but will not close issue experienced by the customer was confirmed and the root cause was determined to be a broken solder joint.The supplier is currently working on improvements to create a more robust soldering process to prevent solder joints from breaking.The instructions for use state the following in regards to product inspection: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Become familiar with the amount of handle movement required to operate the cups.If any irregularities are noted, do not use.Note: exercising the handle while the forceps is coiled may result in damage to the performance characteristics of the forceps." prior to distribution, all captura biopsy forceps are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
During an esophagogastroduodenoscopy (egd), the physician used a captura serrated large forceps-spike.They took one bite.When they went back down for another sample, the forceps would open but would not close.At that point, they felt something snap in the handle.
|