STRYKER SPINE-SWITZERLAND XIA 3 TITANIUM POLYAXIAL SCREW 8.5 X 80MM; PEDICLE SCREW SPINAL SYSTEM Back to Search Results
Model Number 482318580
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 06/09/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that; during final tightening of the blocker to the rod/screw construct the surgeon felt a loosening of the construct.Upon further examination it was noted that the tulip head became dissociated from the screw shaft.The screw was removed easily and a new screw was placed and final tightened without any issues or delays.
 
Event or Problem Description
It was reported that; during final tightening of the blocker to the rod/screw construct the surgeon felt a loosening of the construct.Upon further examination it was noted that the tulip head became dissociated from the screw shaft.The screw was removed easily and a new screw was placed and final tightened without any issues or delays.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM POLYAXIAL SCREW 8.5 X 80MM
Common Device Name
PEDICLE SCREW SPINAL SYSTEM
Manufacturer (Section D)
STRYKER SPINE-SWITZERLAND
le crêt-du-locle 10 a
-
la chaux-de-fonds 2300
CH  2300
MDR Report Key6681582
Report Number3005525032-2017-00076
Device Sequence Number19236322
Product Code MNH
UDI-Device Identifier04546540598653
UDI-Public(01)04546540598653
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K083393
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number482318580
Device Catalogue Number482318580
Device Lot NumberB58631
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2017
Initial Date Received by Manufacturer 06/09/2017
Supplement Date Received by Manufacturer06/09/2017
Initial Report FDA Received Date06/30/2017
Supplement Report FDA Received Date10/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age75 YR
-
-