| Model Number |
48811112 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; 12 mm plate spring bar was deformed while surgeon was implanting.Back up 12 mm plate was used.
|
| |
|
Additional Manufacturer Narrative
|
|
The aviator plate deformation was confirmed via visual inspection.The plate was confirmed to have a spring bar deformed.No relevant manufacturing issues were identified as all units met stryker specifications.The plausible root cause of the event was determined to be using a freehand technique during holes preparation.It is recommended to use variable or fixed drill guide to ensure the proper angulation of the screw so as not to damage the spring bar.
|
| |
|
Event or Problem Description
|
|
It was reported that; 12 mm plate spring bar was deformed while surgeon was implanting.Back up 12 mm plate was used.
|
| |
|