• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 840
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during use, an 840 ventilator had a leak.The patient was transferred to another ventilator.There was no harm to the patient as a result of the event.A service engineer (se) inspected the device and found diagnostic codes relevant to the reported issue but was unable to duplicate the reported condition.The unit passed all testing and operated within the manufacturing specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
MDR Report Key6682407
Report Number8020893-2017-06812
Device Sequence Number1444738
Product Code CBK
UDI-Device Identifier10884521077201
UDI-Public10884521077201
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K970460
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number840
Device Catalogue Number4-840220NISB-EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/09/2017
Initial Report FDA Received Date06/30/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-