AV-TEMECULA-CT XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 1074350-38
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/15/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.Evaluation summary: the device was returned for analysis.The reported material deformation, the reported unstable stent and the reported stent dislodgement were able to be confirmed.The reported difficult to position and the reported difficulty to remove were unable to be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily tortuous and heavily calcified mid right coronary artery.A 3.5 x 38 mm xience xpedition stent delivery system (sds) was being advanced through a guideliner when the stent got stuck and was damaged.The sds was pulled back through the guideliner, the stent became loose and when it was pulled into the co-pilot, it dislodged.The stent was easily removed.A second 3.5 x 38 mm xience xpedition was being advanced when it met resistance with the anatomy and could not cross the lesion.The device was removed and a non-abbott stent was successfully deployed to complete the procedure.There was no clinically significant delay in procedure and no adverse patient effects.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6684596
Report Number2024168-2017-05602
Device Sequence Number150386
Product Code NIQ
UDI-Device Identifier08717648162589
UDI-Public(01)08717648162589(17)191011(10)6100541
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 07/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/11/2019
Device Model Number1074350-38
Device Catalogue Number1074350-38
Device Lot Number6100541
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2017
Initial Report FDA Received Date07/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
-
-