MAUDE Adverse Event Report: SEPTODONT, INC. BIOROOT RCS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SEPTODONT, INC. BIOROOT RCS Back to Search Results
Model Number 01E0300
Medical Device Problem Code Extrusion (2934)
Health Effect - Clinical Code Tingling (2171)
Date of Event 05/06/2017
Type of Reportable Event Serious Injury
Event or Problem Description
(b)(4).Initial information received on (b)(6) 2017 and additional information received on 11-may-2017 and (b)(6) 2017: on (b)(6) 2017 the dentist reported that he performed a root canal procedure on 2nd premolar on mandibular side on a patient with the medical device bioroot.He reported that the medical device extruded reaching the mandibular nerve and leading to lip paresthesia.On 11-may-2017, the dentist reported additional information.The patient was a (b)(6) male, with no specific medical history, who had already been treated by this dentist.The dentist confirmed that he performed dental procedure on tooth #45.He performed a troncular injection of 1/2 cartridge of scandonest 30mg/ml (mepivacaine hydrochloride) (batch number unknown) for anesthesia but this product was not suspected in this report.He used sodium hipochloride 5/25 as root canal irrigant.He performed an x-ray to check : cone was in place, and no extrusion was observed.The patient experienced lip paresthesia, so the dentist performed a scan.An extrusion of the device was observed.The device extruded and reached the external branch of the mandibular nerve.On (b)(6) 2017, the dentist reported some additional information.The patient went to see a maxillo-facial surgeon.The dentist was supposed to see the patient after that.Additional information was expected.Additional information received on 18-may-2017 by phone.The patient was referred to a maxilo-facial surgeon on (b)(6) 2017 for bioroot cement extrusion by removal.The surgery was performed under generalized anesthesia.No information on the patient's status after surgery was provided.At the time of this report, the patient's outcome was unknown.Further information was expected.Additional information received on 08-jun-2017.The dentist saw the patient on (b)(6) 2017 and performed a new scan to check what the surgeon did about product extrusion.The product was taken off the apex but was still present around the nerve (white stripe visible on the scan).The dentist asked the patient to go see the surgeon once again to know why he did not remove the whole product.The patient was still suffering from lip paraesthesia.Causality assessment on 19-may-2017 on initial information received on (b)(6) 2017 and additional information received on 11-may-2017, (b)(6) 2017 and 18-may-2017: seriousness: serious (other medically important condition and required intervention to prevent permanent impairment/damage).Expectedness: paresthesia oral: unlisted eu, unexpected fr; device extrusion: unlisted eu, unexpected fr.Causality: latency: possible.Recognized association: no.Analysis: lip paraesthesia may be due to dental procedure which may had caused direct nerve trauma or to chemical toxicity of irrigation solution.Lip paraesthesia may also be due to compression of the nerve caused by hematoma, by oedema following the dental procedure or by the overfilling spreading of bioroot.In this case, the dentist reports an extrusion of bioroot.The scanner confirms the compression of the external branch of the mandibular nerve.Therefore, the causal relationship between bioroot and paraesthesia is related to the dental procedure with extrusion of the cement and lead to nerve compression.Dechallenge: na.Rechallenge: na.Concluded causality who: probable.Causality assessment on 13-jun-2017 on additional information received on 08-jun-2017: seriousness: serious (other medically important condition and required intervention to prevent permanent impairment/damage).Expectedness: paresthesia oral: unlisted eu, unexpected fr; device extrusion: unlisted eu, unexpected fr.Causality: latency: possible.Recognized association: no.Analysis: the patient still suffers from the paresthesia of the lip because the paste is still present around the branch of the mandibular nerve confirmed by the scanner performed by the dentist.Therefore, the previous assessment applies.Dechallenge: na.Rechallenge: na.Concluded causality who: probable.
 
Additional Manufacturer Narrative
In view no device was returned for evaluation, the production record for the reported suspect device batch was reviewed; no issue identified.The reported batch number was distributed to the following countries: (b)(4).No other complaint has been received for the reported batch number.
 
Event or Problem Description
Additional information received on (b)(4) 2017: the dentist reported the batch number of bioroot was (b)(4).He stated that he has used bioroot several times before on several patients.The dentist saw the patient again on (b)(6) 2017.He reported that the surgeon did not remove the product which was close around the nerve as he thought it was not necessary.The patient had recovered from numbness of lips and chin.Causality assessment on (b)(4) 2017 on additional information received on (b)(4) 2017: seriousness: serious (other medically important condition and required intervention to prevent permanent impairment/damage).Expectedness: paresthesia oral: unlisted eu, unexpected fr, device extrusion: unlisted eu, unexpected fr, causality: latency: possible.Recognized association: no.Analysis: lip paraesthesia may be due to dental procedure considering possible direct nerve trauma, chemical toxicity of irrigation solution or compression of the nerve caused by hematoma, oedema or overfilling spreading of bioroot.In this case, the dentist reported bioroot extrusion, and compression of the external branch of the mandibular nerve was confirmed by scan.According to the last information, the product was removed from the apex by the surgeon but was still present around the nerve.The patient recovered from numbness of lips and chin on unknown date.Therefore this may have play a role in patient recovery.However, as the product has not been completely removed, paraesthesia may be related to the dental procedure with extrusion of the cement which led to nerve compression.Therefore, the causal relationship between bioroot and paraesthesia was assessed as possible, dechallenge: na.Rechallenge: na.Concluded causality who: possible.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOROOT RCS
Common Device Name
BIOROOT RCS
Manufacturer (Section D)
SEPTODONT, INC.
416 s. taylor avenue
louisville CO 80270
MDR Report Key6685396
Report Number0001721729-2017-00006
Device Sequence Number772523
Product Code KIF
UDI-Device Identifier03760140703373
UDI-Public03760140703373
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K130601
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 08/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Model Number01E0300
Device Lot NumberB18997
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2017
Supplement Date Received by Manufacturer06/14/2017
Initial Report FDA Received Date07/04/2017
Supplement Report FDA Received Date08/02/2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age23 YR
-
-