During implantation procedure, noise was noted when the lead was connected to the header of the device, all electrical measurements were good connected to patient system analyzer (psa).Therefore, the device was exchanged and the patient was in stable condition.
|
Analysis of the atrial, right ventricular, and left ventricular connectors on the device revealed no anomalies.Electrical testing showed no indication of conductor fractures or internal shorts.The results of the investigation concluded the analysis of the device was normal with no anomalies found.Based on the information received, the cause of the reported incident could not be conclusively determined.
|