Model Number 8800000
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 03/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No radiographs were received to confirm the event.No product has been returned for investigation.No root cause can be determined at this time.Labeling review: potential adverse events and complications: ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation." warnings, cautions and precautions: ".Confirm heights of screws match with patient's lordosis when securing lock screws.Failure to verify screw height following insertion may result in inadequate rod normalization." ".Final-tighten all lock screws with the counter-torque and torque t-handle.Do not final-tighten through other instruments in the set, as the rod may not be able to normalize to the tulip.Ensure there is enough rod overhang when final tightening, and do not lock down on the conical portion of the rod.Failure to do so may lead to improper lock down of the construct.It may be necessary for the surgeon to add length to the rod, depending upon patient anatomy and desired lordosis." device not returned.
 
Event or Problem Description
On (b)(6) 2017, a revision surgery occurred due to back out of the s1 screw on the left side.The s1 screw was left in situ and the left side of the construct was extended to the s2ai level.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE POLYAXIAL SPINAL SCREWS
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk blvd
san diego CA 92121
MDR Report Key6686376
Report Number2031966-2017-00082
Device Sequence Number150379
Product Code NKB
UDI-Device Identifier00887517344076
UDI-Public00887517344076
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121619
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model Number8800000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2017
Initial Report FDA Received Date07/05/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-