No radiographs were received to confirm the event.No product has been returned for investigation.No root cause can be determined at this time.Labeling review: potential adverse events and complications ".Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s) loss of fixation " warnings, cautions and precautions ".Confirm heights of screws match with patient's lordosis when securing lock screws.Failure to verify screw height following insertion may result in inadequate rod normalization." ".Final-tighten all lock screws with the counter-torque and torque t-handle.Do not final-tighten through other instruments in the set, as the rod may not be able to normalize to the tulip.Ensure there is enough rod overhang when final tightening, and do not lock down on the conical portion of the rod.Failure to do so may lead to improper lock down of the construct.It may be necessary for the surgeon to add length to the rod, depending upon patient anatomy and desired lordosis." device not returned.
|
|
On (b)(6) 2017, a revision surgery occurred due to back out of the left side s1 level screw.The s1 screw was left in situ and the construct was extended.It was reported that patient underwent a subsequent surgery on (b)(6) 2017, due to back out of the s1 screw on the right side.The s1 screw was left in situ and the construct was extended and another screw was placed on the left side.No patient injury was reported.
|