• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY 2CM PERIPHERAL CUTTING BALLOON®; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY 2CM PERIPHERAL CUTTING BALLOON®; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number M001PCB5020900
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device was not received for analysis.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
 
Event Description
It was reported that the balloon broke off the shaft.A 2cm peripheral cutting balloon was selected for use.During procedure, it was noticed that the shaft was in the patient's body and the balloon was still inside.However, the balloon came out of the patient on an unspecified wire.During device withdrawal, it was further noticed that the balloon broke off the catheter's shaft.The procedure was completed with this device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2CM PERIPHERAL CUTTING BALLOON®
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6687587
MDR Text Key79137538
Report Number2134265-2017-06451
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070951
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM001PCB5020900
Device Catalogue NumberPCB502090
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-