| Catalog Number |
2152112 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product is not returned to manufacturer for evaluation however x-ray images are available which revealed the following : "post op x-rays for l5-s1 anterior lumbar interbody fusion show grade ii spondylolisthesis and anterior displacement of l5-s1 cage.We don't have intra-op films to see if this was placed correctly to begin with,but given significant spondylolisthesis suspect that reduction had not occurred at the time of surgery.Given spinal instability and history of parkinson ,posterior augmentation may have been appropriate.".
|
| |
|
Event or Problem Description
|
|
It was reported that during a follow up visit after three months of the operation, surgeon observed that the plate had pulled away from the sacral body.Screws were pulled out of the bone and are still retained in the plate.However, the patient is clinically doing very well and so at this time no further action will be taken.
|
| |
|