| Model Number |
119116 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that, upon inspection of the sample received, there appeared to be excess silicone on the bifurcation valve base and on the tip of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported issue was confirmed.The device was returned for evaluation.Per visual inspection a burr was noted at catheter tip and funnel, no other defects were found.Per dimensional evaluation the flash on molded funnel was found without specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: ¿visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, upon inspection of the sample received, there appeared to be excess silicone on the bifurcation valve base and on the tip of the catheter.
|
| |
|