• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. APOLLO WAND; GWG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. APOLLO WAND; GWG Back to Search Results
Catalog Number AP9
Medical Device Problem Codes Mechanical Problem (1384); Device Operates Differently Than Expected (2913); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
During preparation for a micro neurosurgery, the hospital staff noticed that the apollo wand (wand) was not working properly upon attaching it to the apollo generator and testing the apollo system.It was reported that the wire within the wand was not activated when depressing the apollo generator foot pedal (foot pedal) on the apollo system.Therefore, the physician opened a new wand, tested it and used it to complete the procedure.
 
Additional Manufacturer Narrative
Results: the apollo wand (wand) was leaking where the tubing connects to the rotating hemostasis valve (rhv).Conclusions: evaluation of the returned device revealed that the wire inside the wand was activated when depressing the foot pedal on the apollo system.The wand was connected to an apollo system and powered on.The foot pedal to the apollo generator was depressed and vibration was observed.Therefore, the reported lack of vibration could not be confirmed.Further evaluation of the wand revealed a leak.During functional analysis, while attempting to aspirate and irrigate the wand, two leaks were observed where the tubings are connected to the rotating hemostasis valve (rhv).The root cause of these leaks could not be determined.Penumbra wands are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO WAND
Common Device Name
GWG
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6691748
Report Number3005168196-2017-01097
Device Sequence Number18843200
Product Code GWG
UDI-Device Identifier00814548016344
UDI-Public00814548016344
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K152699
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2018
Device Catalogue NumberAP9
Device Lot NumberF73737
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2017
Initial Date Received by Manufacturer 06/07/2017
Supplement Date Received by Manufacturer07/12/2017
Initial Report FDA Received Date07/06/2017
Supplement Report FDA Received Date08/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-