| Model Number |
2406 |
| Medical Device Problem Code |
Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/12/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Multiple attempts for product return and requests for additional information were made.The product has not been returned to masimo to allow an analysis to be performed.If new information is obtained or the product is returned, a follow up report will be submitted., additional manufacturing narrative (if other): initial reporter zip code exceeded maximum allowable digits, zip code is as follows: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported sensor's led turned off without providing an error message.No patient impact or consequences were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: other, additional manufacturing narrative (if other): the returned cable was evaluated.During inspection, the cable passed visual analysis and software testing; however, failed functional analysis due to an open connection in the cable on the green conductor along the length of the cable.A ¿sensor off patient" error message displayed; which would have prevented the unit from being able to engage in monitoring activities.
|
| |
|