Model Number 1095060
Medical Device Problem Code Maintenance Does Not Comply To Manufacturers Recommendations (2974)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Date of Event 06/16/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the caregiver that the filter of this device was very dirty as the patient had never changed it, and as a result the patient was hospitalised with what the caregiver described as 'bacteria buildup'.It also subsequently transpired that the nebuliser cups and masks had not been being cleaned regularly - but replaced monthly instead.The ifu for the compressor states 'the filter is to be replaced when it appears discolored or wet' and the ifu for the nebuliser states 'after each use, disconnect tubing from sidestream and wash all parts separately in mild liquid detergent and warm water.Rinse thoroughly and shake out excess water.Allow to air dry and reassemble.' replacement filters were provided, and the cleaning requirements explained to the caregiver.The patient has retained the device, and no further issues have been reported with it.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ESSENCE
Common Device Name
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
MDR Report Key6692811
Report Number9681154-2017-00009
Device Sequence Number10955407
Product Code CAF
UDI-Device Identifier00383730001814
UDI-Public00383730001814
Combination Product (Y/N)N
PMA/510(K) Number
K042655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1095060
Device Catalogue Number1099966
Device Lot Number161014
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2017
Initial Report FDA Received Date07/07/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age74 YR
-
-