Model Number R SERIES |
Device Problem
Improper Alarm (2951)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib fault 72" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
|
|
Manufacturer Narrative
|
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing without duplicating the malfunction.Review of the device history log provided evidence of the customer's report.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
|