| Catalog Number |
K6410 SMN 10445103 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer contacted siemens healthcare diagnostics.The cause for the discordant elevated ft4 result is unknown.The customer stated that the quality control was within specification during testing.Mdr 2517506-2017-00520 and 2517506-2017- 00551 were filed for the same event.Siemens is investigating the incident.
|
| |
|
Event or Problem Description
|
|
A discordant elevated free thyroxine (ft4) result was obtained on a patient sample on the dimension vista 500 system.The result was reported to the physician and questioned.The customer was unable to repeat the sample; therefore, no corrected report was issued.There are no reports of patient intervention or adverse health consequences as a result of the discordant elevated ft4 result.
|
| |
|
Additional Manufacturer Narrative
|
|
The original mdr was filed on 2017-07-07.Analysis of the instrument and instrument data indicate the cause for the discordant elevated free thyroxine (ft4) result is unknown.No sample was returned to siemens for testing.A siemens healthcare diagnostics headquarters support center (hsc) representative evaluated the information and found the elevated free thyroxine value is repeatable, patient specific and discordant with the free triiodothyronine and thyroid stimulating hormone values, alternate methodology and the clinical impression.The instructions for use for the ft4 free thyroxine flex® reagent cartridge, contains the following information: "patient samples may contain heterophilic antibodies that could react in immunoassays to give a falsely elevated or depressed result.This assay has been designed to minimize interference from heterophilic antibodies.Nevertheless, complete elimination of this interference from all patient specimens cannot be guaranteed.A test result that is inconsistent with the clinical picture and patient history should be interpreted with caution.Thyroid autoantibodies in human serum may interfere and cause falsely elevated ft4 results." the device is performing within specifications.No further evaluation of the device is required.
|
| |
|
Search Alerts/Recalls
|