| Model Number |
HEALON 5 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Intraocular Pressure Increased (1937)
|
| Date of Event |
12/07/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Implant and explant dates: if implanted or explanted; give date: not applicable as this is not an implantable device.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2016, during the 1-day postoperative visit, the intraocular pressure (iop) was 31 mmhg.The left eye (os) was treated with an iop reducing medication.The patient returned for the unscheduled postoperative visit on (b)(6) 2016.The intraocular pressure os was 20 mmhg.The adverse event was considered resolved without sequelae and the iop reducing medication was discontinued.The patient has recovered.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the complaint sample was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Retained product evaluation: visual inspection on retained products was performed on parallel batch lot# ub32008.48 samples were investigated.No visible defects were found on the package, cannula or in the solutions.The solution was clear and colorless.All components were included in the products and were without defects.No visible defects on the product boxes.The printing on the carton and labels is correct.Chemical and microbiological tests were performed on the retain samples and the results were within product quality specification.Product deficiency was not found on the retain samples.Manufacturing records review: the manufacturing records for the healon were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|