|
(b)(4).This case was reviewed and investigated according to the manufacturer's policy.Attempts to obtain patient identification was unsuccessful.Attempts to obtain the information were made via email and phone.All reasonably known patient information is included in this report.By report, no damage was observed during prep.An insertion needle was used to shape the tip.This was a calcified vessel and the measurement was meant to be done in a diagonal branch.More resistance was felt than a normal straightforward lad intervention.When trying to retrieve the wire, it was suddenly stuck in the lesion.All portions of the device appeared accounted for after removal, but appeared damaged.The spring coil tip distal to the sensor housing appeared extracted.A 1.2 mm balloon was used to retrieve the wire and both devices were retrieved together.A buddy/safety wire and another balloon was also used in order to be prepared for any potential coronary perforation.Other devices used at the same time as the manufacturer's device were a launcher guiding catheter, standard introducer, nothing relevant to the incident.The device was used one time.A good negative ifr reading was obtained indicating no treatment was needed.No additional pci was performed.Cardiac echo/ultrasound was performed to exclude coronary perforation that was not visible on angiogram.Echo was negative, no pericardial fluid could be seen.The patient stayed an extra day in the hospital.The patient is well and by account has not felt this good for years.Device was returned on 7-5-2017.Visual inspection performed during pre-decontamination observed the wire connector was not received for investigation.The tip coil was stretched and the core wire was damaged.The sensor has residue.The dome is present.The instructions for use (ifu) warns, as with all catheterization procedures, complications may be encountered with the use of the pressure guide wire.Do not flex the proximal (electrical connector) end of the pressure guide wire when not inserted in the connector.Excessive flexing can damage or break the internal components.Never advance, torque or retract a pressure guide wire which meets significant resistance.The ifu also cautions the pressure guide wire should not be advanced if resistance is encountered.The wire should never be forcibly pushed into a vessel.Any time that resistance is encountered, the wire should be withdrawn under fluoroscopic guidance.In some instances, the wire may kink and must be removed.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.There were no nonconforming material reports or deviations noted that would contribute to the reported failure mode.This complaint will be monitored as part of complaint data analysis.
|