Investigation summary: the event unit was returned to applied medical for evaluation.Upon visual inspection, the probe tip was detached and damage was observed on the probe and cautery post.Functional testing could not be performed as the device was returned non-functional.Based on the condition of the returned unit, it is likely that the damage observed was caused by improper handling during use.The instructions for use (ifu) warns the user to "inspect the device for any damage, ensuring there are no unintended sharp edges, rough surfaces or protrusions on the outer surface of the electrosurgical probe which may cause harm.Verify that all parts are in working order.Do not use device if damaged.Discard if the device is bent, nicked, cracked, leaking or otherwise damaged." applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.
|