• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-522NAS
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypoglycemia (1912); Reaction (2414)
Event Date 11/02/2016
Event Type  Injury  
Manufacturer Narrative
The insulin pump was received with currents in spec.The pump passed rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.The pump had minor scratched lcd window, cracked case near display window corners, cracked battery tube threads, cracked reservoir tube lip, reservoir tube cracked and missing end cap sticker, and cracked lcd window.
 
Event Description
Customer reported via phone call that they experienced low blood glucose levels of under 60 mg/dl.The customer was treated for the low blood glucose with injection shots.The customer's current blood glucose was 102 mg/dl.The customer stated that he was asleep and had a horrible insulin reaction.The customer was assisted with troubleshooting and stated that the drive support cap was normal.The product was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6695891
MDR Text Key79351677
Report Number3004209178-2017-61912
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 11/07/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-522NAS
Device Catalogue NumberMMT-522NAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/07/2016
Initial Date FDA Received07/08/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age73 YR
Patient Weight81
-
-