• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - CORK GREENFIELD¿; FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - CORK GREENFIELD¿; FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number M001503000
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2017
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause was unable to be determined.(b)(4).
 
Event Description
It was reported that the filter legs did not completely open.A greenfield filter was deployed in the vena cava.It was noted that the legs of the filter did not open completely and appeared stuck.The filter remains implanted.No patient complications were reported and the patient's condition is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENFIELD¿
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
Manufacturer (Section G)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6697537
MDR Text Key79515645
Report Number2134265-2017-07226
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K901659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Model NumberM001503000
Device Catalogue Number50-300
Device Lot Number20106444
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/19/2017
Initial Date FDA Received07/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-