Model Number X SERIES |
Device Problem
Improper Alarm (2951)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "short detected" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical (b)(4); the malfunction was duplicated and the multi-function gasket and multi-function connector were replaced to remedy the problem.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|