• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION E SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number E SERIES
Medical Device Problem Code Improper Alarm (2951)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that during a routine shift check by a clinician, the device displayed a "cable fault" message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was not returned to zoll medical (b)(4); the device's multi-function cable was removed and returned.The multi-function cable was put through extensive testing without duplicating the malfunction.The cable was scrapped.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E SERIES DEFIBRILLATOR
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key6698657
Report Number1220908-2017-01579
Device Sequence Number992355
Product Code MKJ
UDI-Device Identifier00847946016685
UDI-Public00847946016685
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K042007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberE SERIES
Device Catalogue NumberE SERIES
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/21/2017
Supplement Date Received by Manufacturer06/21/2017
Initial Report FDA Received Date07/10/2017
Supplement Report FDA Received Date08/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-