|
It was reported that during an attempt to deploy the endovascular stent graft in the acute angle of the left upper arm fistula, the endovascular stent graft failed to deploy.Reportedly, another endovascular delivery system was used and the procedure was completed.There was no reported patient injury.
|
|
Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Investigation summary: flair endovascular stent graft was returned for evaluation; however the inner product packaging sleeve and label were not returned.Medical records were not provided.No images or photos were provided.No kinks were noted along the catheter and no anomalies noted to proximal handle.The stent graft was found to be partially deployed with graft extending from distal end of outer catheter.Guidewire patency was tested with out issue.An attempt was made to deploy the stent graft, however the catheter stretched and broke.No further testing was able to be performed due to the catheter break.The investigation is confirmed for failure to deploy as the partially deployed stent was not able to be deployed during evaluation.The definitive root cause could not be determined based upon available information.It is unknown whether patient and/or procedural issues contributed to the reported event.Labeling review: indication for use: flair endovascular stent graft is indicated for use in the treatment at the venous anastomosis of eptfe or other synthetic arteriovenous (av) access grafts.Warnings: the stent graft (implant) cannot be repositioned after total or partial deployment.Once partially or fully deployed, the flair endovascular stent graft cannot be retracted or remounted onto the delivery system.Device removal after deployment can only be done with a surgical approach.Precautions: t he safety and effectiveness of this device have not been established when used around a tight bend in a looped av graft.It is recommended to access the av graft at the venous side of the av graft or at the apex.Careful attention should be paid to ensure the device is appropriately sized to the actual graft diameter, taking into account any change to the stated graft diameter that may have resulted from previous interventions.The appropriate length device(s) should be selected so that the stent graft extends beyond the stenosis into at least 10 mm of the non-diseased graft towards the arterial inflow and into the non-diseased vein approximately 10 mm beyond the stenosis.During deployment of the stent graft, the entire length of the delivery system should be kept as straight as possible in order to mitigate the potential for distal stent graft misplacement.After full stent graft deployment, wait a few seconds to allow for complete device expansion before removing the delivery system over the guidewire.Stent graft dislodgement has been reported during removal of the delivery system; therefore, careful attention should be paid during this portion of the procedure to prevent such occurrences.In case of placement of two stent grafts (overlap), use the same device diameter in both cases.If a flared device is used to overlap, do not deploy the flared end inside the first stent graft.Ensure a minimum 10 mm overlap of the devices.Potential complications and adverse events: stent graft specific events that could be associated with clinical complications include stent graft misplacement, stent graft migration, stent graft fracture, stent graft kinking, insufficient stent graft expansion and stent graft embolism.Delivery system specific events that could be associated with clinical complications include bond joint failures, detachment of parts, incompatibility with accessory devices, premature deployment, inaccurate deployment, failure to deploy, high deployment forces, delivery system kinking, no visibility under fluoroscopy, inability to track to target location, and blood leakage from delivery system (hemostasis).Stent graft size selection: special care must be taken to ensure that the appropriate size flair endovascular stent graft is selected.The stent graft body diameter should be approximately 1 mm larger than the synthetic av access graft diameter.
|
|
It was reported that during an attempt to deploy the endovascular stent graft in the acute angle of the left upper arm fistula, the endovascular stent graft failed to deploy.Reportedly, another endovascular stent graft was used to complete the procedure.There was no reported patient injury.
|