| Catalog Number |
54840005540 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/01/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent transforaminal lumbar interbody fusion (tlif) with screws and cage implantation at level l5-s1.Post-op, the screw implanted at the right s1 broke.A tlif revision was performed to replace the split screw.The tip of the screw still remained in the patient.No patient complication have been reported after this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: visual review confirms screw breakage 5 threads down from the base of the bone screw neck.Visual and optical examination of the area of fracture initiation did not identify a pre-existing surface defect near the area of crack origination that could contribute to crack propagation.Some fracture surface smearing is noted.Microscopic examination of the fracture surface identified a fairly flat fracture surface with ratchet marks near the origin of crack propagation, and gently convex progressive striations through the cross-sectional area of the bone screw, consistent with cyclic fatigue.Dimensional examination of the major and neck diameters of the bone screw confirm relevant bone screw dimensions are within print specification.The above observations are consistent with cyclic fatigue.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|