| Model Number |
SCD13 |
| Medical Device Problem Code |
Device Stops Intermittently (1599)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/19/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident ultrasonic dissector was returned for evaluation.Visual inspection of the disposable hand piece revealed that the device had been used and the static part of the jaw had broken off.The broken piece was returned with the rest of device.The remaining waveguide and the broken piece were inspected under magnification to identify the point of initial contact and fracture.It was concluded that the titanium waveguide fractured during use and eventually broke off.The titanium waveguide may have came in contact with a hard metallic object such as hemostat or retractor as evidenced by the break point and metallic scraping.The user¿s guide for this system warns: contact between the active blade and other metal objects (hemostats, clips, staples, retractors, etc.) may result in unintended damage to tissue and/or device failure.
|
| |
|
Event or Problem Description
|
|
The customer originally reported that the device stopped working during the procedure.Another dissector was opened and used to complete the case.There was no patient injury.The device was returned for evaluation and a piece of the active waveguide was disengaged from the device.The broken piece was returned with the remainder of the dissector.
|
| |
|