| Catalog Number |
K6412 SMN 10445104 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer contacted siemens healthcare diagnostics concerning the impact of biotin on the tsh result.The cause for the discordant depressed tsh result is unknown.The customer stated that the quality control was within specification.Mdr 2517506-2017- 00573 and 2517506-2017-00574 were filed for the same event.Siemens is investigating the incident.
|
| |
|
Event or Problem Description
|
|
A discordant depressed thyroid stimulating hormone (tsh) result was obtained on a patient sample on the dimension vista 1500.The result was not reported to the physician.The same patient sample was tested at an alternate facility with an alternate methodology and a higher result was obtained.There are no reports of patient intervention or adverse health consequences as a result of the discordant depressed tsh result.
|
| |
|
Additional Manufacturer Narrative
|
|
The original mdr was filed 7/11/2017.The tsh assay has been tested for potential interference at biotin levels of 100, 250, 500, and 1200 ng/ml.Interference testing was performed following clinical laboratory standards institute (clsi) guidance at concentrations approximately equivalent to the levels listed in the ifu.Below is the biotin interference of dimension vista tsh: test level: 4.24, units: ¿iu/ml, biotin test level and % interference observed, 1200 ng/ml -99.1%, 500 ng/ml -38.7%.250 ng/ml -11.7%, 100 ng/ml -2.4%.The siemens investigation is continuing.
|
| |
|
Additional Manufacturer Narrative
|
|
Original mdr was filed 7/11/2017.Supplemental report was filed 8/22/2017.Siemens ongoing investigation has identified the following new information regarding potential biotin interference, biotin interference limits are currently incorrectly listed in the instructions for use (ifu) for dimension vista thyroid stimulating hormone (tsh) flex reagent cartridges k6412 non-interference information section.The current biotin non-interference ifu information for tsh indicates that the amount of biotin that causes < or = 10% interference is 500 ng/ml (2050 nmol/l).Customers were informed that concentrations of biotin above 100 ng/ml (409 nmol/l) can potentially result in interference > 10%.Siemens issued an urgent medical device correction (umdc) communication (b)(4) dated october 16, 2017 to all u.S.Dimension vista® accounts and an urgent field safety notice (b)(4), to all outside u.S.Dimension vista® accounts.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr 2517506-2017-00575 was filed 7/11/2017.Mdr supplement 2517506-2017-00575 supplement 1 was filed 8/22/2017.Mdr supplement 2517506-2017-00575 supplement 2 was filed 11/10/2017.Based on input from the (b)(6) recall coordinator on 12/18/2017 this recall was resubmitted to (b)(6) under 2432235-12/19/2017-004-c.Changed from 2517506-03/30/2017-003-c to 2432235-12/19/2017-004-c.
|
| |
|
Search Alerts/Recalls
|