• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE EASYTRAK 2; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE EASYTRAK 2; IMPLANTABLE LEAD Back to Search Results
Model Number 4542
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Endocarditis (1834); Pocket Erosion (2013)
Date of Event 05/10/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event or Problem Description
Boston scientific received information that this left ventricular (lv) lead was explanted due to pocket erosion.It was noted that this patient had a history of endocarditis.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
MDR Report Key6702668
Report Number2124215-2017-08748
Device Sequence Number19172592
Product Code NIK
UDI-Device Identifier00802526579806
UDI-Public(01)00802526579806(17)20170518
Combination Product (Y/N)N
PMA/510(K) Number
P010012/S024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/18/2017
Device Model Number4542
Other Device ID NumberEASYTRAK 2
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/10/2017
Initial Report FDA Received Date07/11/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4542; 7740; 7741; 7742; L301
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age61 YR
-
-