| Catalog Number |
3269.SBS06 |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/12/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device history lot review: unable to perform since the lot number is unknown.Sample analysis: it was determined on 13jun2017 that an examination of the returned light fiber revealed that the coating was melted as a result of direct contact with a hot object.The damaged coating causes the fiber not to function in a proper manner.Additionally, it was identified that the damage most likely originated from the outside rather than from within the fiber itself and strengthens our observation that an external heat source was the cause.The location of the damage was found to be located somewhere halfway between the connector and the hand piece and not anywhere near the patients eye.Conclusion: the complaint has not been found acknowledged.It is unknown when this incident occurred.We are not aware of any injury that occurred.Should any further information be obtained a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
An initial report has been received reporting "upon arriving the office manager at the center handed me a stack of bad product." no further information was reported.The sample for this complaint has been returned and evaluated.
|
| |
|