| Model Number |
11.1510 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
05/06/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The dhr review shows the device performed as intended when it was manufactured.The tens device and handheld remote control were able to power on.The device output treatment frequencies were to specification, however, the waveform was not to specification.The device is will be evaluated by the design engineers for root cause of the waveform not performing to specification.The device labeling was reviewed and found to be suitable and adequate for the device to perform its intended use.
|
| |
|
Event or Problem Description
|
|
The customer complained that when she pressed the power button on the remote control to start the treatment, the device gave her a "jolt" and then stopped working altogether.
|
| |
|