| Catalog Number |
500-55150 |
| Medical Device Problem Codes |
Fracture (1260); Device Inoperable (1663); Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/26/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The catheter (sn: (b)(4)) was returned to ekos on 7/7/17 for evaluation.The catheter evaluation confirmed the ultrasound core was fractured and the fracture occurred at the shaft to treatment zone joint.Examination of the ultrasound core shows indications of a prior kink in the fracture region, and some bending on internal wires.During follow up, ekos qa was informed that the patient anatomy was not tortuous and free of any graft.A definitive root cause could not be identified.Ekosonic endovascular device instructions for use states: do not deform or kink the ultrasound core during delivery into the infusion catheter.If the ultrasound core is kinked at any time, do not attempt to use it, as kinking may lead to degraded performance or fracture during use.Review of the manufacturing records indicates that the product was manufactured according to specification following all applicable procedures and the finished device met all acceptance criteria prior to release.
|
| |
|
Event or Problem Description
|
|
A 106cm/50cm ekosonic catheter (sn: (b)(4)) was used to treat a patient with peripheral arterial occlusion on (b)(6) 2017.After successfully placed infusion catheter (part of the ekosonic catheter) in the patient, the treating physician tried to place ultrasound core (the microsonic device) and the ultrasound core wire fractured inside the infusion catheter.The physician was able to retrieve ultrasound core by pulling out the infusion catheter and replacing with another ekos device.The procedure was completed and patient was doing well.
|
| |
|
Search Alerts/Recalls
|