• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ELEGANCE; COMPRESSOR, AIR, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ELEGANCE; COMPRESSOR, AIR, PORTABLE Back to Search Results
Model Number 1095059
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/22/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the prongs on the plug of this device remained in the mains socket when the patient tried to unplug it.On return of the device for evaluation, it was confirmed that both prongs were missing from the plug.It would therefore appear that a risk of electrocution may have been created by having metal prongs protruding from a live mains socket.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ELEGANCE
Common Device Name
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
MDR Report Key6705126
Report Number9681154-2017-00013
Device Sequence Number10913599
Product Code CAF
UDI-Device Identifier00383730001630
UDI-Public00383730001630
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K042655
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1095059
Device Catalogue Number1100134
Device Lot Number151209
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2017
Initial Report FDA Received Date07/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-