Model Number EX1150
Device Problems Charred (1086); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2017
Event Type  malfunction  
Event Description
Patient¿s caregiver reported that the transmitter failed to power up after the storm.Attempts to slide the prongs off to reset them and trying different outlets did not resolve the issue.It was also noted that there was black charring on the left contact.The transmitter was replaced.
 
Manufacturer Narrative
Final analysis found that upon opening the ac power adapter, burnt components were observed on the circuit board.Upon opening the transmitter, burn damage was noted on multiple modem circuit components.A burn mark on the inner casing of the ac power adapter was also noted.It was concluded that the damage could be due to high current flow through the ac wall power outlet which resulted in the power anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6707579
MDR Text Key79835931
Report Number2017865-2017-05384
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup,Followup
Report Date 08/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Other Device ID Number05414734504799
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2017
Date Manufacturer Received08/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age89 YR
-
-