| Model Number |
CL3534 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lot review: two suspect lots were reported: lot # 3436827 (mfg date 5/2017) shows (b)(4) units and lot # 3432850 (mfg date 4/2017) shows (b)(4) units were manufactured, tested, inspected, and released with no exception documents cited.
|
| |
|
Event or Problem Description
|
|
Chemolock adapter used as part of baxter iv administration set for circle priming and administration of chemotherapy.Chemotherapy infusion was discovered to be leaking, specifically the chemotherapy adapter, approximately 5 minutes into chemotherapy infusion.Product reached patient, no apparent harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot review: two suspect lots were reported: lot # 3436827 (mfg date 5/2017) shows (b)(4) units and lot # 3432850 (mfg date 4/2017) shows (b)(4) units were manufactured, tested, inspected, and released with no exception documents cited.Visual analysis: cl3534 was received with mating device and found to be leaking at chemolock.Functional analysis: returned cl3534 was pressure tested.Leakage was detected at the chemolock spin mechanism.The o-ring was found to be misassembled on the post.Summary: the cl3534 was confirmed to be leaking.The cause was a mis-assembled o-ring.
|
| |
|
Event or Problem Description
|
|
Chemolock adapter used as part of baxter iv administration set for circle priming and administration of chemotherapy.Chemotherapy infusion was discovered to be leaking, specifically the chemotherapy adapter, approximately 5 minutes into chemotherapy infusion.Product reached patient, no apparent harm.
|
| |
|
Search Alerts/Recalls
|