| Model Number |
423714 |
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device was misaligned with the dot indicator used for insertion.As a result, the device could not be inserted.
|
| |
|
Additional Manufacturer Narrative
|
|
Received 1 clean catheter in open packaging.The reported event was unconfirmed since the problem could not be reproduced.Per visual inspection performed on returned sample, the position of notch or dot on the catheter funnel was aligned with catheter tip.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.To use: insert rounded end of catheter." (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the device was misaligned with the dot indicator used for insertion.As a result, the device could not be inserted.
|
| |
|
Search Alerts/Recalls
|