MEDTRONIC MINIMED PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723RNAB
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 06/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call the insulin pump alarmed button error.The customer's blood glucose was 500 mg/dl at the time of incident.Customer stated that the insulin pump has been dropped prior to button error alarm and the buttons were unresponsive.Button error troubleshooting was performed and confirmed that time is advancing.Customer also reported to have experienced a high blood glucose of 500 mg/dl.Customer did not mention any form of treatment for the high blood glucose event and no troubleshooting was performed.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Additional Manufacturer Narrative
Findings: insulin pump received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm during testing noted.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime and excessive no delivery test due to buttons not responding.Pump received with cracked case at the display window corner, cracked reservoir tube lip, minor scratched lcd window, cracked belt clip slot and cracked battery tube threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723RNAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325 1219
MDR Report Key6709498
Report Number2032227-2017-28184
Device Sequence Number992753
Product Code OYC
UDI-Device Identifier00613994748713
UDI-Public(01)00613994748713
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 08/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-723RNAB
Device Catalogue NumberMMT-723RNAB
Device Lot NumberA000167986
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/27/2017
Initial Date Received by Manufacturer 06/17/2017
Supplement Date Received by Manufacturer08/09/2017
Initial Report FDA Received Date07/12/2017
Supplement Report FDA Received Date08/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
Patient Weight210
-
-