Model Number 734028/10ADP
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Codes Fever (1858); Seroma (2069)
Date of Event 06/03/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).A review of manufacturing and qc records confirmed that device has been manufactured to specification.Vascutek did not receive complaints for any other units from the complaint batches or any other units that were impregnated and sterilised at the same time as the complaint device.Endotoxin results for the complaint device were reviewed and they were within the limit.As the event was reported as fever and seroma 5-year similar events review covered all infection, seroma and fever complaints for all gelweave family product types which gave a low occurrence rate of 0.002% (complaints v sales).During review of vascutek's clinical evaluation report and relevant clinical papers for polyester vascular grafts, it was noted that seroma formation is a known and expected complication of this type of procedure.As a complaint device was not available physical analysis can't be completed.Vascutek now consider this event closed; however, the event type will continue to be tracked and trended through the complaint's monitoring system.If negative trend arises corrective actions may be considered at the time.
 
Event or Problem Description
Vascutek have received notification of the event that was described as 39-degree fever one week after the implant day.A seroma around the implanted graft was observed as well.Initially procedure was complete as usual.The patient is in clinical follow-up stage and is on the way to recovery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GELWEAVE VALSALVA ANTE-FLO
Common Device Name
GELWEAVE VALSALVA ANTE-FLO
Manufacturer (Section D)
VASCUTEK LTD
newmains avenue
inchinnan business park
renfrewshire, scotland PA4 9 RR
UK  PA4 9RR
MDR Report Key6710368
Report Number9612515-2017-00011
Device Sequence Number1948341
Product Code DSY
UDI-Device Identifier05037881121031
UDI-Public05037881121031
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K013022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/13/2017
Report Date (Section F) 06/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/30/2021
Device Model Number734028/10ADP
Device Catalogue Number734028/10ADP
Device Lot Number15871214-9326
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Device Age10 MO
Event Location Hospital
Date Report to Manufacturer06/28/2017
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/28/2017
Initial Report FDA Received Date07/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-