MEDTRONIC NEUROSURGERY CSF-FLOW CONTROL VALVE, BUTTON, MEDIUM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
|
Back to Search Results |
|
| Catalog Number |
24003 M |
| Medical Device Problem Codes |
Leak/Splash (1354); Occlusion Within Device (1423)
|
| Health Effect - Clinical Code |
Therapeutic Response, Decreased (2271)
|
| Date of Event |
06/15/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for reflux testing.However, the valve did not meet requirements for pressure-flow and pre-implantation testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the returned valve also did not meet the requirements for leak testing due to a tear in bottom of the valve.It is unknown how or when this damage occurred.The instructions for use (ifu) that accompany the device caution that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, crushing or breaking of components.¿ the ifu also caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.¿ a review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported to medtronic neurosurgery that the valve was not functioning and there was no drainage of cerebrospinal fluid.In addition, there was a suture leak.According to the report, the shunt was completely removed and would be replaced with a different shunt after two weeks.Reportedly, there was no harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for reflux testing.However, the valve did not meet requirements for pressure-flow and pre-implantation testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the returned valve also did not meet the requirements for leak testing due to a tear in bottom of the valve.It is unknown how or when this damage occurred.The instructions for use (ifu) that accompany the device caution that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, crushing or breaking of components.¿ the ifu also caution that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.¿ a review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|