• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY CARDIAC/PERITONEAL CATHETER, STANDARD, BARIUM IMPREGNATED, 90 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY CARDIAC/PERITONEAL CATHETER, STANDARD, BARIUM IMPREGNATED, 90 CM; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Catalog Number 43103
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Therapeutic Response, Decreased (2271)
Date of Event 06/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Approximately 89 cm of the peritoneal catheter was returned.The returned catheter was patent and met the requirements for leak testing.A review of the manufacturing records showed no anomalies.All catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported to medtronic neurosurgery that the valve was not functioning and there was no drainage of cerebrospinal fluid.In ad dition, there was a suture leak.According to the report, the shunt was completely removed and would be replaced with a different shunt after two weeks.Reportedly, there was no harm or injury.
 
Additional Manufacturer Narrative
Approximately 89 cm of the peritoneal catheter was returned.The returned catheter was patent and met the requirements for leak testing.A review of the manufacturing records showed no anomalies.All catheters are 100% inspected at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIAC/PERITONEAL CATHETER, STANDARD, BARIUM IMPREGNATED, 90 CM
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key6715529
Report Number2021898-2017-00379
Device Sequence Number16455477
Product Code JXG
UDI-Device Identifier00643169466821
UDI-Public00643169466821
Combination Product (Y/N)N
Initial Reporter CountryTS
PMA/510(K) Number
K792005
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2020
Device Catalogue Number43103
Device Lot NumberE01045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2017
Supplement Date Received by Manufacturer07/28/2017
Initial Report FDA Received Date07/14/2017
Supplement Report FDA Received Date08/21/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age1 MO
Patient Weight4
-
-