|
The returned valve was patent.The valve met requirements for reflux, pressure-flow, and preimplantation testing.There was crystalline debris observed within the interior and exterior of the valve.The instructions for use that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ the returned valve did not meet the requirements for leak testing due a tear observed in the top of the proximal occluder.It is unknown how or when this damage occurred.The ifu cautions that ¿care should be taken in handling the valves as silicone has a low cut and tear resistance.¿ the ifu also cautions that ¿improper use of instruments in the handling or implantation of shunt products may result in the cutting, slitting, crushing or breaking of components.¿ a review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
It was reported to medtronic neurosurgery that the device had a hole in the cerebrospinal fluid (csf) flow output and was found intr a-operatively.According to the report, the doctor used another valve to complete the surgery.Reportedly, there was no injury, and the patient was stable.
|