STRYKER SPINE-FRANCE AVIATOR ASSY TWO LEVEL PLATE SIZE 30; SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS. Back to Search Results
Model Number 48811230
Medical Device Problem Codes Material Deformation (2976); Material Integrity Problem (2978)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/22/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after the surgeon performed a two level acdf on a patient he happen to touch the plate prior to closing the incision and realized there was something sharp protruding.He then noticed that the locking spring ring had popped out.
 
Additional Manufacturer Narrative
Manufacturing files were reviewed and no anomalies were found.The plate was confirmed visually to have a deformed spring bar.The plausible root cause is undetermined.
 
Event or Problem Description
It was reported that after the surgeon performed a two level acdf on a patient he happen to touch the plate prior to closing the incision and realized there was something sharp protruding.He then noticed that the locking spring ring had popped out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVIATOR ASSY TWO LEVEL PLATE SIZE 30
Common Device Name
SPINAL INTERVERTEBRAL BODY FIXATION ORTHOSIS.
Manufacturer (Section D)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR  33610
MDR Report Key6720066
Report Number0009617544-2017-00287
Device Sequence Number16455458
Product Code KWQ
UDI-Device Identifier07613252152194
UDI-Public(01)07613252152194
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number48811230
Device Catalogue Number48811230
Device Lot Number166630
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2017
Initial Date Received by Manufacturer 06/22/2017
Supplement Date Received by Manufacturer06/22/2017
Initial Report FDA Received Date07/17/2017
Supplement Report FDA Received Date10/25/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No