| Model Number |
48811230 |
| Medical Device Problem Codes |
Material Deformation (2976); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that after the surgeon performed a two level acdf on a patient he happen to touch the plate prior to closing the incision and realized there was something sharp protruding.He then noticed that the locking spring ring had popped out.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing files were reviewed and no anomalies were found.The plate was confirmed visually to have a deformed spring bar.The plausible root cause is undetermined.
|
| |
|
Event or Problem Description
|
|
It was reported that after the surgeon performed a two level acdf on a patient he happen to touch the plate prior to closing the incision and realized there was something sharp protruding.He then noticed that the locking spring ring had popped out.
|
| |
|