| Model Number |
MMT-7008B |
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/21/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's father reported via phone call that his daughter experienced sensor glucose versus blood glucose differences with threshold suspend.Customer's sensor glucose was 40 and blood glucose was 450 mg/dl.Customer was advised that their blood glucose and sensor glucose were not within acceptable range.The caller declined troubleshooting for any other issues.The insulin pump will not be returning for analysis.The sensor will be returning for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened/used sensor and performed continuity resistance test found sensor passed per specifications and performed the sensor initialization test using a new lab transmitter with a 7xx/5xx paradigm insulin pump.Connected the sensor to the transmitter and placed it in 100 bts solution, confirmed the communication icon was displayed and that the transmitter was flashing while connected to the sensor.The sensor functioned properly.Found cannula bent unable to confirm customer received sensor in said condition due to customer return sensors opened/used.Cannot performed bts test as the sensor is beyond its expiration date.
|
| |
|